To: | Sysmex Corporation (officeactions@brinksgilson.com) |
Subject: | U.S. Trademark Application Serial No. 88907077 - XR - 11333-1028 |
Sent: | November 30, 2020 04:19:58 PM |
Sent As: | ecom115@uspto.gov |
Attachments: |
United States Patent and Trademark Office (USPTO)
Office Action (Official Letter) About Applicant’s Trademark Application
U.S. Application Serial No. 88907077
Mark: XR
|
|
Correspondence Address:
|
|
Applicant: Sysmex Corporation
|
|
Correspondence Email Address: |
|
NON-FINAL OFFICE ACTION
The USPTO must receive applicant’s response to this letter within six months of the issue date below or the application will be abandoned. Respond using the Trademark Electronic Application System (TEAS). A link to the appropriate TEAS response form appears at the end of this office action.
Issue date: November 30, 2020
This office action is in response to applicant’s communication filed on November 13, 2020.
Applicant Must Amend The Identification
Applicant must clarify some of the wording in the identification, as shown below, because it remains indefinite and overly broad. See 37 C.F.R. §2.32(a)(6); TMEP §§1402.01, 1402.03. The wording is indefinite because it does not make clear the exact nature of the goods. Further, it is overly broad because it encompasses goods in more than one international class. Specifically, certain goods are properly classified in Class 9 if for laboratory use and Class 10 if for clinical use. Additionally, applicant should avoid the use of a namely clause within a namely clause because it introduces a lack a clarity.
The wording in the suggested identification that appears in bold and italics shows the additions that are being proposed. Wording that
appears within brackets offers guidance. And wording that should be deleted is shown with a line through it as follows: strikethrough. Applicant should enter amendments in standard font, not
in bold, italics or strikethrough or within brackets. Please note that wording that appears in strikethrough in the suggested identification that is copied and pasted into a response will not be
deleted but rather will appear in standard font.
Applicant may adopt the following identification, if accurate:
“Instruments and apparatus for genetic testing, namely, for DNA analysis or detection of abnormal cell, for laboratory use; medical instruments and apparatus for gene testing, namely, polymerase chain reaction analyzer for laboratory use; sample transfer apparatus specially adapted for transferring containers of blood samples for laboratory use, sample container transfer apparatus specially adapted for transferring containers of blood samples for laboratory use, sample rack transfer apparatus specially adapted for transferring racks of containers of blood samples for laboratory use,” in International Class 9; and
“Diagnostic or measuring apparatus and instruments for medical purposes, namely, apparatus for use in blood analysis, medical instruments and apparatus for analyzing
reticulocytes, medical instruments and apparatus for analyzing erythrocytes, medical instruments and apparatus for analyzing blood coagulation, medical instruments and apparatus for analyzing
platelet aggregation, and bodily fluids measuring apparatus and instruments in the nature of devices for measuring or analysing white blood cells and red blood cells; blood collecting devices,
namely, blood sampling tube, holders specially adapted for blood sampling tubes, dropping pipettes for medical purposes and needles for medical purposes sold empty; apparatus for blood analysis; blood testing apparatus; hematology analyzers for medical diagnostic
uses; hematology analyzers, namely, blood diluting apparatus for medical diagnostic use; blood cell analyzers for medical purposes; blood pretreatment apparatus and instruments, namely, {specify what the goods are}; cell counters, namely, apparatus for counting live or dead cells for diagnosis; blood cell counters, namely, apparatus for
counting red blood cell, white blood cell or platelet for diagnosis; platelet counters, namely, apparatus for counting platelets for diagnosis; particle counters, namely, apparatus for counting
particles in a liquid for medical purposes; diagnostic or measuring apparatus and instruments for medical purposes, namely, medical instruments and apparatus for analyzing urine, medical instruments
and apparatus for analyzing urine cells, medical instruments and apparatus for analyzing urinary particles, medical instruments and apparatus for urine pretreatment, namely, in the nature of apparatus for use in medical analysis of urine, medical instruments and apparatus for analyzing immunochemicals, immunological testing apparatus
for medical use, medical apparatus and instruments in the nature of immunoassay analyzers, medical apparatus and instruments for clinical chemistry analyzing, namely, for the biochemical
analysis of bodily fluids, medical instruments and apparatus for analyzing occult blood, medical instruments and apparatus for analyzing bacteria, medical instruments and apparatus for genetic
testing, namely, in particular, for DNA analysis or detection of abnormal cell, for clinical use, medical instruments and apparatus for gene testing, namely, in the nature of a polymerase chain
reaction analyzer for clinical use; sample transfer apparatus specially adapted for transferring containers of
blood samples for clinical use, sample container transfer apparatus specially adapted for transferring
containers of blood samples for clinical use, sample rack transfer apparatus specially adapted for transferring
racks of containers of blood samples for clinical use; furniture especially made for medical purposes; furniture especially made for medical purposes,
namely, table units specifically adapted for use with medical apparatus, medical supplies, medical equipment and instruments; furniture especially made for medical purposes, namely, instrument
stands, carts and cabinets, all for hospital use; furniture especially made for medical purposes, namely, stands, carts and cabinets specifically adapted for use with medical equipment, medical
supplies and instruments; furniture especially made for medical purposes, namely, mobile workbenches with storage cabinets, specifically adapted for placing analyzers thereon for hospital use;
structural and structural fitted replacement parts for all the aforementioned goods,” in International Class 10.
The requirement for an acceptable, amended identification is continued and maintained.
For assistance with identification and classification in trademark applications, please consult the USPTO’s online searchable ID Manual. See TMEP §1402.04.
Multi-Class Application Requirements
(1) List the goods by their international class number in consecutive numerical order, starting with the lowest numbered class.
(2) Submit a filing fee for each international class not covered by the fee already paid (view the USPTO’s current fee schedule). The application identifies goods that are classified in at least two classes. However, applicant submitted a fee sufficient for only one class. Applicant must either submit the filing fees for the classes not covered by the submitted fees or restrict the application to the number of classes covered by the fees already paid.
See 37 C.F.R. §2.86(a); TMEP §§1403.01, 1403.02(c).
For an overview of the requirements for a Section 1(b) multiple-class application and how to satisfy the requirements online using the Trademark Electronic Application System (TEAS) form, see the multiple-class application webpage.
Requirement For Information
This requirement applies to (i) each item in the identification that requires amendment for which applicant adopts wording that is not identical to the suggested identification, and (ii) blood pretreatment apparatus and instruments.
To permit proper examination of the application, applicant must submit additional product information about applicant’s goods because the nature of such goods is not clear from the present record. See 37 C.F.R. §2.61(b); TMEP §814. The requested product information should include fact sheets, instruction manuals, advertisements and promotional materials, and/or a photograph of the identified goods. If these materials are unavailable, applicant should submit similar documentation for goods of the same type, explaining how its own product will differ. Applicant must also describe in detail the nature, purpose, and channels of trade of the goods.
Factual information about the goods must clearly indicate how they operate, their salient features, and their prospective customers and channels of trade.
Failure to comply with a request for information is grounds for refusing registration. In re Harley, 119 USPQ2d 1755, 1757-58 (TTAB 2016); TMEP §814. Merely stating that information about the goods is available on applicant’s website, or submitting weblinks without attaching the relevant pages, is an insufficient response and will not make the relevant information of record. See In re Planalytics, Inc., 70 USPQ2d 1453, 1457-58 (TTAB 2004).
Applicant is invited to contact the assigned examining attorney with any questions regarding this action.
/Katherine S. Chang/
Trademark Examining Attorney
Law Office 115
571-270-1528
katherine.chang@uspto.gov
RESPONSE GUIDANCE